全自动比表面积及孔径分析仪英文介绍

日期:2017-09-11 16:59:34/ 人气: / 来源:

Rise-1001 Automatic specific surface area and Pore size distribution Analyzer

Brief Introduction to the Instrument
      Rise-1001 Automatic specific surface area and Pore size distribution Analyzer accords with the international standards ISO9277, ISO 15901 and National standard Gb-119587. It tests desorption process in the principle of static capacity measurement tested through static gas balance and pressure tested. They are tested in the temperature of liquid nitrogen. The gas of known quantity is filled into the sample tube and the pressure is to fall, then  the Moore's quality of  adsorbed gas is calculated. With calculated balance adsorption quantity and theoretical model, the instrument evaluates the parameters such as  single-point, multipoint  BET specific surface area, Langmuir specific surface area of the samples; BJH mesoporous,     macroporous area,area distribution; t-plot microporous and surface areas, Dubinin Astakhov microporous distribution,  Hovath-Kawazoe microporous distribution; DFT and MC aperture distribution model.
Guarantee for Trustworthy Tested Data
      Reasonable connection of pipes, capability of s vacuum pumping, precision and sensitivity of sensors, automation of the instrument and numbers of the digital models for the different samples.
Modular Gas pipes Design
      Rise-1001 Automatic specific surface area and Pore size distribution Analyzer accords with the most developed design ideas, designed with special modular stainless steel vacuum pipes . It is leak proof and pollution-free. It is high vacuum and avoids leakage due to excessive pipe connections. It is specially-designed research instrument for the micropores study, with diffusion pump to keep the system high vacuum.
Constant temperature
      Dewar flasks is self-developed with special internal structure, store the liquid nitrogen for more than 10 days. The liquid nitrogen is almost free of losses during the test so that the temperature of the tested samples is constant and that the Dewar flasks are protected as they are fragile and unfavorable for movement.
High-precision sensors
      Several high-precision sensors and 22-bit AD converters ensure the precision of the specific surface area and aperture.
Developed theoretical model
      With DFT and MC aperture distribution models, it is developed in the world, and is the Number One of the specific surface area and Pore size distribution Analyzers in China
Automatic operation
      Adsorption and desorption processes are automatically controlled by computers and free of manual operation.
Technical characteristics
  1、 single-point, multipoint BET specific surface area
  2、 BJH specific surface area, area distribution, total area
  3、t-plot micropores volume and surface areas, Dubinin Astakhov micropores size distribution, Hovath-Kawazoe micropores size distribution
  4、 DFT and MC aperture distribution model
Technical parameters and instrument configuration
 1、 Test principle: low-temperature nitrogen physical adsorption (static capacity)
 2、 specific surface area : 0.0005m 2 without upper limit
 3、 Analysis of range for pore size : 3.5-5000  angstrom
 4、 adsorbed gas: nitrogen
 5、 P/PO test range: 1 X 10 -8 – 0.995
 6、 pressure and temperature measurement:
      2 imported absolute pressure sensors 0-120 Kpa( precision 0.1%)
      1 imported pressure sensors 0-1 Kpa( precision 0.1%)
      1 imported pressure sensors 0-100pa( precision 0.1%)
      1 temperature sensor PT-100( Precision 0.1)
 7、 Dewar flask: 2.0L, duration of 80 hours
 8、 vacuum pump:  Mechanical pump + diffusion pump
 9、 Ultimate vacuum: 1.0 X 10 -6 Torr
 10、 Measurement software:  adsorption /desorption isotherm curve
 11、 1tested sample and 3 pretreated samples   in the meantime
Output Report
      Directly printed and Excel output the parameters such as adsorption /desorption isotherm, BET specific surface area, Langmuir specific surface area , t-plot micropores areas, BJH specific surface area ,Dubinin Astakhov micropores distribution,  Hovath-Kawazoe micropores distribution.
Applied range
      It is applied in study and test of all kinds of materials such as zeolite, carbon material, molecular sieve, silicic acid anhydride, aluminum oxide,soil, clay, catalytic agent, organic metal  compound cageconstruction.